CVE-2017-12720: An Improper Access Control issue was discovered in Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pum
CVE-2017-12720 is a high-severity flaw (CVSS 8.1) affecting medfusion 4000 wireless syringe infusion pump. Below: why it ranks where it does, which versions are affected, what to do, and the exact commands.
An Improper Access Control issue was discovered in Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump, Version 1.1, 1.5, and 1.6. The FTP server on the pump does not require authentication if the pump is configured to allow FTP connections.
Why this priority
- High severity
- Remotely exploitable, no login, no user action
No fix is published yet: follow the downgrade or mitigation steps.
Risk facts
| CVSS score | 8.1 (High) |
| Attack vector | Network (remote) |
| Attack complexity | High |
| Privileges needed | None |
| User interaction | None |
| EPSS exploit chance | 1.9% (top 21%) |
| CISA exploited list | No |
| Weakness | CWE-306 |
| Published | 2018-02-15 |
| Last modified | 2026-06-17 |
Last updated 2026-06-17 · Sources: CVE Program, NVD, CISA KEV, EPSS and vendor advisories · How we score
Am I affected?
Timeline
- Published CVE record published
- Updated Record last modified
Scores and fixes here are computed automatically from those and other public sources. Confirm with the vendor advisory before changing production systems.
Quick answers
How do I fix CVE-2017-12720?
medfusion 4000 wireless syringe infusion pump: mitigate or apply the listed mitigation. Exact commands for each OS are on this page.
Is CVE-2017-12720 being exploited?
CVE-2017-12720 is not on the CISA Known Exploited Vulnerabilities list at the time of the last update. Check again, as this can change.
What does CVE-2017-12720 affect?
CVE-2017-12720 affects medfusion 4000 wireless syringe infusion pump. Severity is high (CVSS 8.1). An Improper Access Control issue was discovered in Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump, Version 1.1, 1.5, and 1.6. The FTP server on the pump does not require authentication if the pump is configured to allow FTP co
How urgent is CVE-2017-12720?
Priority 53/100. High severity; Remotely exploitable, no login, no user action
Related CVEs
- CVE-2017-12726 HIGH 7.3 MITIGATE A Use of Hard-coded Password issue was discovered in Smiths Medical Medfusion 4000 Wireless Syringe Infusion P
- CVE-2017-12725 MEDIUM 5.6 MITIGATE A Use of Hard-coded Credentials issue was discovered in Smiths Medical Medfusion 4000 Wireless Syringe Infusio
- CVE-2017-12724 HIGH 8.1 MITIGATE A Use of Hard-coded Credentials issue was discovered in Smiths Medical Medfusion 4000 Wireless Syringe Infusio
- CVE-2017-12723 LOW 3.7 MITIGATE A Password in Configuration File issue was discovered in Smiths Medical Medfusion 4000 Wireless Syringe Infusi
- CVE-2017-12722 MEDIUM 5.3 MITIGATE An Out-of-bounds Read issue was discovered in Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump, Ve