Free and open · new CVEs checked every 20 minutesStatusRSSOpen dataContact

CVE-2019-5024: A restricted environment escape vulnerability exists in the “kiosk mode” function of Capsule Technologies Smar

37
Routine
HIGH 7.6UPGRADE

CVE-2019-5024 is a high-severity flaw (CVSS 7.6) affecting smartlinx neuron 2 firmware. Below: why it ranks where it does, which versions are affected, what to do, and the exact commands.

A restricted environment escape vulnerability exists in the “kiosk mode” function of Capsule Technologies SmartLinx Neuron 2 medical information collection devices running versions 9.0.3 or lower. A specific series of keyboard inputs can escape the restricted environment, resulting in full administrator access to the underlying operating system. An attacker can connect to the device via USB port with a keyboard or other HID device to trigger this

Why this priority

  • High severity

A fixed release or distro patch exists.

Risk facts

CVSS score7.6 (High)
Attack vectorPhysical
Attack complexityLow
Privileges neededNone
User interactionNone
EPSS exploit chance0.5% (top 59%)
CISA exploited listNo
WeaknessCWE-693
Published2019-04-11
Last modified2026-06-17
Publisher (CNA): Talos · Vendor: capsuletech

Share: LinkedIn · X · Reddit

Last updated 2026-06-17 · Sources: CVE Program, NVD, CISA KEV, EPSS and vendor advisories · How we score

Am I affected?

Timeline

capsuletech smartlinx neuron 2 firmware UPGRADE
Affected: <= 9.0.3
Fixed in: newer than 9.0.3
Safe to downgrade to: no unaffected older release
Check your OS release
cat /etc/os-release; uname -r
Debian / Ubuntu
sudo apt update && sudo apt full-upgrade
RHEL / Amazon Linux / Fedora
sudo dnf upgrade --refresh
Windows
Install the latest cumulative update from Windows Update
If no patch exists yet
OS downgrades are not safe. Apply the vendor mitigation from the advisory and restrict exposure.
Verify at the official source NVD · CVE Program · Report a problem with this page
Scores and fixes here are computed automatically from those and other public sources. Confirm with the vendor advisory before changing production systems.

Quick answers

How do I fix CVE-2019-5024?

smartlinx neuron 2 firmware: upgrade to newer than 9.0.3. Exact commands for each OS are on this page.

Is CVE-2019-5024 being exploited?

CVE-2019-5024 is not on the CISA Known Exploited Vulnerabilities list at the time of the last update. Check again, as this can change.

What does CVE-2019-5024 affect?

CVE-2019-5024 affects smartlinx neuron 2 firmware. Severity is high (CVSS 7.6). A restricted environment escape vulnerability exists in the “kiosk mode” function of Capsule Technologies SmartLinx Neuron 2 medical information collection devices running versions 9.0.3 or lower. A specific series of keyboard inputs can es

How urgent is CVE-2019-5024?

Priority 37/100. High severity