CVE-2020-15486: An issue was discovered on Dr Trust ECG Pen 2.00.08 devices

34
Routine
MEDIUM 6.5MITIGATE

CVE-2020-15486 is a medium-severity flaw (CVSS 6.5) affecting electrocardiogram pen firmware. Below: why it ranks where it does, which versions are affected, what to do, and the exact commands.

An issue was discovered on Dr Trust ECG Pen 2.00.08 devices. Because the Bluetooth LE support is implemented without a requirement for pairing or security, any attacker can access the GATT server of the device and can sniff the data being broadcasted while a measurement is being done. Also, saved data can also be extracted over a Bluetooth connection. In addition, an attacker can launch a man-in-the-middle attack against data integrity.

Why this priority

  • No elevated risk signals yet

No fix is published yet: follow the downgrade or mitigation steps.

Risk facts

CVSS score6.5 (Medium)
Attack vectorAdjacent network
Attack complexityLow
Privileges neededNone
User interactionNone
EPSS exploit chance0.5% (top 57%)
CISA exploited listNo
Published2020-08-26
Last modified2026-06-17
Publisher (CNA): MITRE · Vendor: drtrust

Share: LinkedIn · X · Reddit

Last updated 2026-06-17 · Sources: CVE Program, NVD, CISA KEV, EPSS and vendor advisories · How we score

Am I affected?

Timeline

drtrust electrocardiogram pen firmware MITIGATE
Affected: 2.00.08
Fixed in: no fixed release known
Safe to downgrade to: a release before 2.00.08
Check your OS release
cat /etc/os-release; uname -r
Debian / Ubuntu
sudo apt update && sudo apt full-upgrade
RHEL / Amazon Linux / Fedora
sudo dnf upgrade --refresh
Windows
Install the latest cumulative update from Windows Update
If no patch exists yet
OS downgrades are not safe. Apply the vendor mitigation from the advisory and restrict exposure.
Verify at the official source NVD · CVE Program · Report a problem with this page
Scores and fixes here are computed automatically from those and other public sources. Confirm with the vendor advisory before changing production systems.

Quick answers

How do I fix CVE-2020-15486?

electrocardiogram pen firmware: mitigate or apply the listed mitigation. Exact commands for each OS are on this page.

Is CVE-2020-15486 being exploited?

CVE-2020-15486 is not on the CISA Known Exploited Vulnerabilities list at the time of the last update. Check again, as this can change.

What does CVE-2020-15486 affect?

CVE-2020-15486 affects electrocardiogram pen firmware. Severity is medium (CVSS 6.5). An issue was discovered on Dr Trust ECG Pen 2.00.08 devices. Because the Bluetooth LE support is implemented without a requirement for pairing or security, any attacker can access the GATT server of the device and can sniff the data being b

How urgent is CVE-2020-15486?

Priority 34/100. No elevated risk signals yet.