CVE-2020-6966: In ApexPro Telemetry Server Versions 4.2 and prior, CARESCAPE Telemetry Server v4.2 & prior, Clinical Informat
CVE-2020-6966 is a critical-severity flaw (CVSS 10.0) affecting apexpro telemetry server firmware, carescape central station mai700 firmware, carescape central station mas700 firmware. Below: why it ranks where it does, which versions are affected, what to do, and the exact commands.
In ApexPro Telemetry Server Versions 4.2 and prior, CARESCAPE Telemetry Server v4.2 & prior, Clinical Information Center (CIC) Versions 4.X and 5.X, CARESCAPE Central Station (CSCS) Versions 1.X, the affected products utilize a weak encryption scheme for remote desktop control, which may allow an attacker to obtain remote code execution of devices on the network.
Why this priority
- Critical severity
- Remotely exploitable, no login, no user action
A fixed release or distro patch exists.
Risk facts
| CVSS score | 10.0 (Critical) |
| Attack vector | Network (remote) |
| Attack complexity | Low |
| Privileges needed | None |
| User interaction | None |
| EPSS exploit chance | 2.2% (top 17%) |
| CISA exploited list | No |
| Weakness | CWE-326 |
| Published | 2020-01-24 |
| Last modified | 2026-06-17 |
Last updated 2026-06-17 · Sources: CVE Program, NVD, CISA KEV, EPSS and vendor advisories · How we score
Am I affected?
Timeline
- Published CVE record published
- Updated Record last modified
Scores and fixes here are computed automatically from those and other public sources. Confirm with the vendor advisory before changing production systems.
Quick answers
How do I fix CVE-2020-6966?
apexpro telemetry server firmware: upgrade to newer than 4.2; carescape central station mai700 firmware: mitigate or apply the listed mitigation; carescape central station mas700 firmware: mitigate or apply the listed mitigation. Exact commands for each OS are on this page.
Is CVE-2020-6966 being exploited?
CVE-2020-6966 is not on the CISA Known Exploited Vulnerabilities list at the time of the last update. Check again, as this can change.
What does CVE-2020-6966 affect?
CVE-2020-6966 affects apexpro telemetry server firmware, carescape central station mai700 firmware, carescape central station mas700 firmware. Severity is critical (CVSS 10.0). In ApexPro Telemetry Server Versions 4.2 and prior, CARESCAPE Telemetry Server v4.2 & prior, Clinical Information Center (CIC) Versions 4.X and 5.X, CARESCAPE Central Station (CSCS) Versions 1.X, the affected products utilize a weak encrypt
How urgent is CVE-2020-6966?
Priority 61/100. Critical severity; Remotely exploitable, no login, no user action
Software pages
Related CVEs
- CVE-2020-6965 CRITICAL 9.9 UPGRADE MITIGATE In ApexPro Telemetry Server Versions 4.2 and prior, CARESCAPE Telemetry Server v4.2 & prior, Clinical Informat
- CVE-2020-6964 HIGH 8.6 UPGRADE MITIGATE In ApexPro Telemetry Server Versions 4.2 and prior, CARESCAPE Telemetry Server v4.2 & prior, Clinical Informat
- CVE-2020-6963 CRITICAL 10.0 UPGRADE MITIGATE In ApexPro Telemetry Server Versions 4.2 and prior, CARESCAPE Telemetry Server v4.2 & prior, Clinical Informat
- CVE-2020-6962 CRITICAL 10.0 UPGRADE MITIGATE In ApexPro Telemetry Server, Versions 4.2 and prior, CARESCAPE Telemetry Server v4.2 & prior, Clinical Informa