CVE-2015-3954: Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and S
CVE-2015-3954 is a critical-severity flaw (CVSS 9.8) affecting plum a\+ infusion system firmware, plum a\+3 infusion system firmware, symbiq infusion system firmware. Below: why it ranks where it does, which versions are affected, what to do, and the exact commands.
Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and Symbiq Infusion System, version 3.13 and prior give unauthenticated users root privileges on Port 23/TELNET by default. An unauthorized user could issue commands to the pump. Hospira recommends that customers close Port 23/TELNET on the affected devices. Hospira has also released the Plum 360 Infusion System which is not vulnerable to this
Why this priority
- Critical severity
- Remotely exploitable, no login, no user action
A fixed release or distro patch exists.
Risk facts
| CVSS score | 9.8 (Critical) |
| Attack vector | Network (remote) |
| Attack complexity | Low |
| Privileges needed | None |
| User interaction | None |
| EPSS exploit chance | 2.0% (top 19%) |
| CISA exploited list | No |
| Weakness | CWE-285 |
| Published | 2019-03-25 |
| Last modified | 2026-06-17 |
Last updated 2026-06-17 · Sources: CVE Program, NVD, CISA KEV, EPSS and vendor advisories · How we score
Am I affected?
Timeline
- Published CVE record published
- Updated Record last modified
Scores and fixes here are computed automatically from those and other public sources. Confirm with the vendor advisory before changing production systems.
Quick answers
How do I fix CVE-2015-3954?
plum a\+ infusion system firmware: upgrade to newer than 13.4; plum a\+3 infusion system firmware: upgrade to newer than 13.6; symbiq infusion system firmware: upgrade to newer than 3.13. Exact commands for each OS are on this page.
Is CVE-2015-3954 being exploited?
CVE-2015-3954 is not on the CISA Known Exploited Vulnerabilities list at the time of the last update. Check again, as this can change.
What does CVE-2015-3954 affect?
CVE-2015-3954 affects plum a\+ infusion system firmware, plum a\+3 infusion system firmware, symbiq infusion system firmware. Severity is critical (CVSS 9.8). Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and Symbiq Infusion System, version 3.13 and prior give unauthenticated users root privileges on Port 23/TELNET by default. An unauthor
How urgent is CVE-2015-3954?
Priority 60/100. Critical severity; Remotely exploitable, no login, no user action
Related CVEs
- CVE-2015-3956 CRITICAL 9.8 UPGRADE UPGRADE Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and S
- CVE-2015-3953 CRITICAL 9.8 UPGRADE UPGRADE Hard-coded accounts may be used to access Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Inf
- CVE-2015-3952 HIGH 7.5 UPGRADE UPGRADE Wireless keys are stored in plain text on Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Inf