CVE-2015-3956: Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and S
CVE-2015-3956 is a critical-severity flaw (CVSS 9.8) affecting plum a\+ infusion system firmware, plum a\+3 infusion system firmware, symbiq infusion system firmware. Below: why it ranks where it does, which versions are affected, what to do, and the exact commands.
Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and Symbiq Infusion System, version 3.13 and prior accept drug libraries, firmware updates, pump commands, and unauthorized configuration changes from unauthenticated devices on the host network. Hospira recommends that customers close Port 20/FTP and Port 23/TELNET on the affected devices. Hospira has also released the Plum 360 Infusion Syste
Why this priority
- Critical severity
- Remotely exploitable, no login, no user action
A fixed release or distro patch exists.
Risk facts
| CVSS score | 9.8 (Critical) |
| Attack vector | Network (remote) |
| Attack complexity | Low |
| Privileges needed | None |
| User interaction | None |
| EPSS exploit chance | 1.0% (top 39%) |
| CISA exploited list | No |
| Weakness | CWE-345 |
| Published | 2019-03-25 |
| Last modified | 2026-06-17 |
Last updated 2026-06-17 · Sources: CVE Program, NVD, CISA KEV, EPSS and vendor advisories · How we score
Am I affected?
Timeline
- Published CVE record published
- Updated Record last modified
Scores and fixes here are computed automatically from those and other public sources. Confirm with the vendor advisory before changing production systems.
Quick answers
How do I fix CVE-2015-3956?
plum a\+ infusion system firmware: upgrade to newer than 13.4; plum a\+3 infusion system firmware: upgrade to newer than 13.6; symbiq infusion system firmware: upgrade to newer than 3.13. Exact commands for each OS are on this page.
Is CVE-2015-3956 being exploited?
CVE-2015-3956 is not on the CISA Known Exploited Vulnerabilities list at the time of the last update. Check again, as this can change.
What does CVE-2015-3956 affect?
CVE-2015-3956 affects plum a\+ infusion system firmware, plum a\+3 infusion system firmware, symbiq infusion system firmware. Severity is critical (CVSS 9.8). Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and Symbiq Infusion System, version 3.13 and prior accept drug libraries, firmware updates, pump commands, and unauthorized configurati
How urgent is CVE-2015-3956?
Priority 55/100. Critical severity; Remotely exploitable, no login, no user action
Related CVEs
- CVE-2015-3954 CRITICAL 9.8 UPGRADE UPGRADE Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and S
- CVE-2015-3953 CRITICAL 9.8 UPGRADE UPGRADE Hard-coded accounts may be used to access Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Inf
- CVE-2015-3952 HIGH 7.5 UPGRADE UPGRADE Wireless keys are stored in plain text on Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Inf