Free and open · new CVEs checked every 20 minutesStatusRSSOpen dataContact

CVE-2015-3956: Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and S

55
This month
CRITICAL 9.8REMOTEUPGRADEUPGRADEUPGRADE

CVE-2015-3956 is a critical-severity flaw (CVSS 9.8) affecting plum a\+ infusion system firmware, plum a\+3 infusion system firmware, symbiq infusion system firmware. Below: why it ranks where it does, which versions are affected, what to do, and the exact commands.

Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and Symbiq Infusion System, version 3.13 and prior accept drug libraries, firmware updates, pump commands, and unauthorized configuration changes from unauthenticated devices on the host network. Hospira recommends that customers close Port 20/FTP and Port 23/TELNET on the affected devices. Hospira has also released the Plum 360 Infusion Syste

Why this priority

  • Critical severity
  • Remotely exploitable, no login, no user action

A fixed release or distro patch exists.

Risk facts

CVSS score9.8 (Critical)
Attack vectorNetwork (remote)
Attack complexityLow
Privileges neededNone
User interactionNone
EPSS exploit chance1.0% (top 39%)
CISA exploited listNo
WeaknessCWE-345
Published2019-03-25
Last modified2026-06-17
Publisher (CNA): ICS-CERT · Vendor: pifzer

Share: LinkedIn · X · Reddit

Last updated 2026-06-17 · Sources: CVE Program, NVD, CISA KEV, EPSS and vendor advisories · How we score

Am I affected?

Timeline

pifzer plum a\+ infusion system firmware UPGRADE
Affected: <= 13.4
Fixed in: newer than 13.4
Safe to downgrade to: no unaffected older release
Check your OS release
cat /etc/os-release; uname -r
Debian / Ubuntu
sudo apt update && sudo apt full-upgrade
RHEL / Amazon Linux / Fedora
sudo dnf upgrade --refresh
Windows
Install the latest cumulative update from Windows Update
If no patch exists yet
OS downgrades are not safe. Apply the vendor mitigation from the advisory and restrict exposure.
pifzer plum a\+3 infusion system firmware UPGRADE
Affected: <= 13.6
Fixed in: newer than 13.6
Safe to downgrade to: no unaffected older release
Check your OS release
cat /etc/os-release; uname -r
Debian / Ubuntu
sudo apt update && sudo apt full-upgrade
RHEL / Amazon Linux / Fedora
sudo dnf upgrade --refresh
Windows
Install the latest cumulative update from Windows Update
If no patch exists yet
OS downgrades are not safe. Apply the vendor mitigation from the advisory and restrict exposure.
pifzer symbiq infusion system firmware UPGRADE
Affected: <= 3.13
Fixed in: newer than 3.13
Safe to downgrade to: no unaffected older release
Check your OS release
cat /etc/os-release; uname -r
Debian / Ubuntu
sudo apt update && sudo apt full-upgrade
RHEL / Amazon Linux / Fedora
sudo dnf upgrade --refresh
Windows
Install the latest cumulative update from Windows Update
If no patch exists yet
OS downgrades are not safe. Apply the vendor mitigation from the advisory and restrict exposure.
Verify at the official source NVD · CVE Program · Report a problem with this page
Scores and fixes here are computed automatically from those and other public sources. Confirm with the vendor advisory before changing production systems.

Quick answers

How do I fix CVE-2015-3956?

plum a\+ infusion system firmware: upgrade to newer than 13.4; plum a\+3 infusion system firmware: upgrade to newer than 13.6; symbiq infusion system firmware: upgrade to newer than 3.13. Exact commands for each OS are on this page.

Is CVE-2015-3956 being exploited?

CVE-2015-3956 is not on the CISA Known Exploited Vulnerabilities list at the time of the last update. Check again, as this can change.

What does CVE-2015-3956 affect?

CVE-2015-3956 affects plum a\+ infusion system firmware, plum a\+3 infusion system firmware, symbiq infusion system firmware. Severity is critical (CVSS 9.8). Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and Symbiq Infusion System, version 3.13 and prior accept drug libraries, firmware updates, pump commands, and unauthorized configurati

How urgent is CVE-2015-3956?

Priority 55/100. Critical severity; Remotely exploitable, no login, no user action

Related CVEs

References

← Back to latest CVEs

Get the weekly exploited-CVE digest by emailThe CVEs being attacked this week and how to fix them. No spam, unsubscribe any time.