Free and open · new CVEs checked every 20 minutesStatusRSSOpen dataContact

CVE-2025-53963: An issue was discovered on Thermo Fisher Ion Torrent OneTouch 2 INS1005527 devices

49
This month
CRITICAL 9.8REMOTEMITIGATE

CVE-2025-53963 is a critical-severity flaw (CVSS 9.8) affecting ion torrent onetouch 2 firmware. Below: why it ranks where it does, which versions are affected, what to do, and the exact commands.

An issue was discovered on Thermo Fisher Ion Torrent OneTouch 2 INS1005527 devices. They run an SSH server accessible over the default port 22. The root account has a weak default password of ionadmin, and a password change policy for the root account is not enforced. Thus, an attacker with network connectivity can achieve root code execution. NOTE: This vulnerability only affects products that are no longer supported by the maintainer.

Why this priority

  • Critical severity
  • Remotely exploitable, no login, no user action

No fix is published yet: follow the downgrade or mitigation steps.

Risk facts

CVSS score9.8 (Critical)
Attack vectorNetwork (remote)
Attack complexityLow
Privileges neededNone
User interactionNone
EPSS exploit chance0.5% (top 60%)
CISA exploited listNo
WeaknessCWE-521
Published2025-12-04
Last modified2026-06-17
Publisher (CNA): MITRE · Vendor: thermofisher

Share: LinkedIn · X · Reddit

Last updated 2026-06-17 · Sources: CVE Program, NVD, CISA KEV, EPSS and vendor advisories · How we score

Am I affected?

Timeline

thermofisher ion torrent onetouch 2 firmware MITIGATE
Affected: all versions
Fixed in: no fixed release known
Safe to downgrade to: no unaffected older release
Check your OS release
cat /etc/os-release; uname -r
Debian / Ubuntu
sudo apt update && sudo apt full-upgrade
RHEL / Amazon Linux / Fedora
sudo dnf upgrade --refresh
Windows
Install the latest cumulative update from Windows Update
If no patch exists yet
OS downgrades are not safe. Apply the vendor mitigation from the advisory and restrict exposure.
Verify at the official source NVD · CVE Program · Report a problem with this page
Scores and fixes here are computed automatically from those and other public sources. Confirm with the vendor advisory before changing production systems.

Quick answers

How do I fix CVE-2025-53963?

ion torrent onetouch 2 firmware: mitigate or apply the listed mitigation. Exact commands for each OS are on this page.

Is CVE-2025-53963 being exploited?

CVE-2025-53963 is not on the CISA Known Exploited Vulnerabilities list at the time of the last update. Check again, as this can change.

What does CVE-2025-53963 affect?

CVE-2025-53963 affects ion torrent onetouch 2 firmware. Severity is critical (CVSS 9.8). An issue was discovered on Thermo Fisher Ion Torrent OneTouch 2 INS1005527 devices. They run an SSH server accessible over the default port 22. The root account has a weak default password of ionadmin, and a password change policy for the r

How urgent is CVE-2025-53963?

Priority 49/100. Critical severity; Remotely exploitable, no login, no user action

Related CVEs

References

← Back to latest CVEs

Get the weekly exploited-CVE digest by emailThe CVEs being attacked this week and how to fix them. No spam, unsubscribe any time.