bd vulnerabilities
Recent CVEs affecting bd products. 17 tracked.
bd has 17 tracked vulnerabilities, 0 critical. A fixed release or vendor patch exists for 9. Most recent: 2023-11-28. Highest priority: CVE-2023-30563 (score 38/100).
- CVE-2023-30565 LOW 3.5 UPGRADE An insecure connection between Systems Manager and CQI Reporter application could expose infusion data to an a
- CVE-2023-30564 MEDIUM 6.9 UPGRADE Alaris Systems Manager does not perform input validation during the Device Import Function.
- CVE-2023-30563 HIGH 8.2 UPGRADE A malicious file could be uploaded into a System Manager User Import Function resulting in a hijacked session.
- CVE-2023-30562 MEDIUM 6.7 UPGRADE A GRE dataset file within Systems Manager can be tampered with and distributed to PCUs
- CVE-2023-30561 MEDIUM 6.1 UPGRADE The data flowing between the PCU and its modules is insecure
- CVE-2023-30560 MEDIUM 6.8 UPGRADE The configuration from the PCU can be modified without authentication using physical connection to the PCU
- CVE-2023-30559 MEDIUM 5.2 UPGRADE The firmware update package for the wireless card is not properly signed and can be modified.
- CVE-2023-29066 LOW 3.2 DOWNGRADE The FACSChorus software does not properly assign data access privileges for operating system user accounts
- CVE-2023-29065 MEDIUM 4.1 DOWNGRADE The FACSChorus software database can be accessed directly with the privileges of the currently logged-in user
- CVE-2023-29064 MEDIUM 4.1 DOWNGRADE The FACSChorus software contains sensitive information stored in plaintext
- CVE-2023-29063 LOW 2.4 DOWNGRADE The FACSChorus workstation does not prevent physical access to its PCI express (PCIe) slots, which could allow
- CVE-2023-29062 LOW 3.8 DOWNGRADE The Operating System hosting the FACSChorus application is configured to allow transmission of hashed user cre
- CVE-2023-29061 MEDIUM 5.2 DOWNGRADE There is no BIOS password on the FACSChorus workstation
- CVE-2023-29060 MEDIUM 5.4 DOWNGRADE The FACSChorus workstation operating system does not restrict what devices can interact with its USB ports
- CVE-2022-47376 HIGH 7.3 UPGRADE The Alaris Infusion Central software, versions 1.1 to 1.3.2, may contain a recoverable password after the inst
- CVE-2022-43557 MEDIUM 5.3 MITIGATE MITIGATE The BD BodyGuard™ infusion pumps specified allow for access through the RS-232 (serial) port interface
- CVE-2022-40263 MEDIUM 6.6 UPGRADE BD Totalys MultiProcessor, versions 1.70 and earlier, contain hardcoded credentials
FAQ
How do I fix bd vulnerabilities?
Open the CVE for your version, check the affected range, then follow the listed action: upgrade to the fixed release, or downgrade or mitigate if no fix exists yet. Commands for each OS are shown on every CVE page.
Is bd being exploited right now?
None of the tracked bd CVEs are on the CISA Known Exploited Vulnerabilities list at the moment. This is checked every 20 minutes.
Which bd CVE should I patch first?
CVE-2023-30563 currently ranks highest (38/100): A malicious file could be uploaded into a System Manager User Import Function resulting in a hijacked session..
Products
- alaris 8015 pcu firmware
- alaris guardrails editor
- alaris infusion central
- alaris systems manager
- bodyguard 323 colorvision firmware
- bodyguard 999-603 firmware
- bodyguard duo 999-903 firmware
- bodyguard epidural 999-683 firmware
- bodyguard pain manager 999-803 firmware
- bodyguard t 999-103 firmware
- facschorus
- guardrails cqi reporter
- totalys multiprocessor firmware